RELIABLE BY DESIGN
ENS-4810DC
Merk:INLANX
Beschrijving: ENS-4810DC Acculaad- en ontlaadtester
Weerstand, nominaal vermogen, tolerantie, TCR, behuizing, draadlengte, montagemethode, markering en verpakking kunnen allemaal worden aangepast.
Productseries op de site kunnen als alternatief dienen voor EBG, HVR, Vishay en andere merken. Klanten moeten monsters en specificaties controleren voordat zij vervangende modellen op grote schaal gebruiken.
Totale voorraad:100
- Grootte
- 460*260*360 mm
- Serie
- ENS-4810DC Acculaad- en ontlaadtester
- Merk
- INLANX
Beschrijving: ENS-4810DC Acculaad- en ontlaadtester
Totale voorraad:100
Merk:INLANX
Productgegevens
Productbeschrijving
For battery-group charge/discharge, activation and maintenance testing; records group and cell voltages, charge/discharge current, time and capacity to assess energy-storage performance and load capacity.
Technische kenmerken
- Load source: custom nickel-chromium alloy resistors with low resistance and high power density; resistance accuracy can be controlled within ±0.001 Ω; low temperature coefficient and resistance to high-current surges.
- Intelligent control: adjusts the load as battery voltage changes to maintain constant-current discharge; acquires cell voltages in real time, displays curves and analyzes voltage status, with multiple configurable assessment thresholds.
- Discharge testing: supports constant-current or constant-power discharge through the internal load with the battery group offline, or constant-current discharge in parallel with user equipment. Voltage, current, time and capacity thresholds are configurable; discharge parameters can be changed during testing.
- Discharge termination: stops testing at the group cutoff voltage, capacity or time limit, when any cell voltage falls below its configured lower limit, or on manual termination; records group voltage, cell voltages, current, time and discharged capacity.
- Automatic charging: uses three stages, constant current → constant voltage with tapering current (equalization voltage regulation) → automatic transition to trickle/float charging, for battery charging recovery and maintenance.
- Touch operation: 7-inch high-brightness TFT resistive touchscreen with 1024×600 resolution and direct on-screen operation.
- Wireless cell monitoring (optional): LoRa modules support 2V, 4V, 6V and 12V cells; each module monitors 6 cells simultaneously. The source example is a 48V battery group fitted with 4 modules.
- Cell-data analysis: displays cell voltages in real time during charging and discharging, supports bar charts and data tables, and automatically identifies the highest- and lowest-voltage cells to help analyze trends.
- Curves and reports: allows review of battery-group voltage and current curves; supports USB-drive data transfer, with analysis software parsing data and generating reports.
Technische specificaties
| Discharge voltage | 10-60 VDC |
|---|---|
| Discharge current | 0-100 A; maximum discharge power 6 kW |
| Charge voltage | 10-60 VDC |
| Charge current | 0-100 A; maximum charge power 6 kW |
| Dimensions (length × width × height) | 460 × 260 × 360 mm |
| Main-unit mass | Approx. 21.5 kg |
| AC supply | Single-phase AC220V, 40-60Hz |
| Operation | Touchscreen |
| Display | 7-inch TFT LCD, resistive touch, 1024×600 |
| Data communication | RS485×1 |
| Internal storage | 128 Mbit FLASH |
| Voltage measurement accuracy | ±0.5%FS + 0.1 V |
| Current measurement accuracy | ±1%FS + 0.1 A |
| Group voltage display accuracy | 0.1 V |
| Group current display accuracy | 0.1 A |
| Discharge current control accuracy | ±1%FS |
| Main-unit protection | Shutdown on overtemperature, overcurrent or loss of current control |
| Shutdown device | High-voltage DC switch |
| Reverse-polarity protection | Supported |
| Abnormal-condition protection | Power loss at the power cable or main cable |
| Overtemperature thresholds | Resistor enclosure 85°C; heatsink 100°C |
| Alarm | LCD indication and buzzer |
| Cooling | Forced-air cooling |
| Operating temperature | -5 to +50°C |
| Storage temperature | -40 to +70°C |
| Relative humidity | 0-90% RH (40±2°C) |
| Rated altitude | 2000 m |
Toepassingen en selectie
- Suitable for battery and UPS power-supply maintenance.